Ten Changes the EU's MDR and IVDR Revision Would Make, Ranked by Who Gains

Share:
Brussels is rewriting its device rules again. We ranked the ten broadest changes still alive in the Commission proposal and the Parliament draft by who actually benefits, who is objecting, and what exporters in India and Asia-Pacific should do now

Ten months after the European Commission published its targeted revision of the Medical Devices Regulation and the In Vitro Diagnostic Regulation on 16 December 2025, the file has reached the point where positions start to harden. On 14 October 2026, EU ambassadors in Coreper I gave broad backing to the Irish Presidency's aim of reaching a Council general approach before its term ends in December. In Parliament, rapporteur Oliver Schenk's draft report for the public health committee (SANT) carries more than 130 amendments, and a plenary vote is expected in Q1 2027. Emergo by UL, writing on 9 October 2026, expects a compromise and adoption around summer 2027, with most simplifications applying six months after entry into force.

Most coverage lists what changes. This ledger asks a sharper question: who gains? And it adds a column nobody in Brussels is writing for, the manufacturers in India, China, Korea and South-East Asia who sell into Europe through an authorised representative and a notified body they rarely meet in person.

How this list was built. A change qualifies if it appears in the Commission proposal (COM(2025) 1023) and remains live in the SANT draft report or in Council discussions as of 23 October 2026. More than ten qualify; we ranked by breadth, scoring each change on how many certified manufacturers it touches (all, one risk segment, or a niche). The proposed deletion of manufacturer liability cover is excluded because the draft report reinstates it. Sources: legal analyses, the Council progress report of 5 June 2026, and published positions from industry, notified bodies, insurers and clinicians.

 

Rank

Change

Biggest winner

Loudest objector

Odds

1

Open-ended certificates

Already-certified manufacturers

Team-NB

Medium

2

Lighter surveillance audits

Firms with clean records

Team-NB

Medium

3

Clocks, dialogue and an ombudsperson

Applicants of every size

Notified bodies

High

4

International reliance chapter

Non-EU exporters

Council legal caution

High

5

Wider clinical evidence and equivalence

Follow-on device makers

Insurers, clinicians

Medium

6

Rule 11 software down-classing

Software developers

German insurers

Medium

7

Fee cuts for micro and small firms

SMEs

Member states, small notified bodies

Medium

8

Well-established technology route

Mature product lines

Insurers

High

9

Orphan, breakthrough and niche pathways

Specialist innovators

Few outright

High

10

Looser in-house IVD rules

Hospital laboratories

MedTech Europe

Medium

 

◆   ◆   ◆

1. Certificates without an expiry date

The proposal scraps the five-year ceiling on notified body certificates. Instead of a full recertification cycle, notified bodies would run periodic, risk-based reviews, possibly folded into annual quality system surveillance. The SANT draft keeps the principle but would hold the initial certificate for niche, orphan and breakthrough devices to five years, renewable only if post-market clinical follow-up conditions are met.

  • Gains: Every manufacturer holding an MDR or IVDR certificate, which is why it tops the ledger. MedTech Europe's 5 May 2026 position backs open validity with risk-based reviews.
  • Loses or worries: Team-NB lists the removal of recertification and expiry dates among changes that would shift oversight "from a preventive to a predominantly reactive model". The Biomedical Alliance objects to unlimited validity for high-risk devices. Notified bodies also lose a predictable revenue line.

Odds of surviving trilogue: Medium. The principle has Commission, industry and rapporteur support, but expect carve-outs for high-risk classes once the Council's safety-minded delegations weigh in.

 

For exporters outside the EU: finish certification under today's rules now; a certificate in hand is the asset this change rewards, and the current MDR remains fully applicable until the revision applies.

2. Fewer, lighter surveillance audits

From certificates to the audits that sit behind them. The Commission would let notified bodies replace on-site audits with remote ones, space surveillance audits up to two years apart where there are no safety concerns, confine unannounced audits to cases with cause, and drop systematic sampling of technical documentation. The draft report ties the 24-month interval to compliance history and places unannounced audits inside a risk-based plan.

  • Gains: Manufacturers with clean audit histories, and anyone far from their notified body.
  • Loses or worries: Team-NB argues the cumulative cuts would push oversight below the level of the old Directives. The Council's 5 June progress report records member state worries about the cumulative effect of simplification on post-market safety.

Odds of surviving trilogue: Medium. Remote audits look safe; the two-year interval will likely survive only with conditions attached, as the rapporteur already proposes.

 

For exporters outside the EU: remote audits cut auditor travel to Asian sites, but a compliance-history test means your next two audits matter more; close open nonconformities now.

3. Clocks, structured dialogue and an ombudsperson

Predictability is the quiet prize. The proposal introduces structured dialogue between manufacturers and notified bodies and an ombudsperson-style mechanism to settle disputes within 90 days. A companion draft implementing regulation, published on 12 December 2025, sets maximum conformity assessment timelines of 255 days, or 165 days where audits and verification run in parallel. The draft report adds pre-submission consultations and aligns companion diagnostic timelines with linked medicine reviews.

  • Gains: Applicants of every size, especially those budgeting launches from abroad.
  • Loses or worries: Team-NB says clock-stopping requirements impose work that is hard to invoice. The Council report questions whether the dispute-settlement mechanism is workable.

Odds of surviving trilogue: High. Structured dialogue is uncontroversial; the ombudsperson is the piece most likely to be reshaped.

 

For exporters outside the EU: ask your notified body for a pre-submission meeting as soon as the text allows, and plan EU launches against the 255-day ceiling rather than hope.

4. A chapter on international reliance

Here the ledger turns to the change that matters most outside Europe. The proposal adds an international cooperation chapter: the Commission could conclude administrative arrangements with third countries, rely on conformity assessments and inspections by foreign regulators, and align with IMDRF and the Medical Device Single Audit Program (MDSAP). The draft report would broaden the use of ISO 13485 and MDSAP audit results.

  • Gains: Non-EU manufacturers already audited by trusted regulators. MedTech Europe wants full EU membership of MDSAP and MDSAP certificates usable for CE marking, noting that Australia has reduced and Brazil has removed recognition of CE marking.
  • Loses or worries: The opposition is procedural rather than political. Most Council delegations support cooperation but insist on Treaty procedures and Council prerogatives, which could slow any actual arrangement.

Odds of surviving trilogue: High. The chapter will survive; the real question is how many arrangements follow, and how fast.

 

For exporters outside the EU: Arnold and Porter frames reliance as a future benefit, not a current right; keep your notified body route intact and build MDSAP audit evidence that maps cleanly to MDR requirements.

5. Wider clinical evidence and easier equivalence

From audits to evidence. The proposal counts published literature and peer-reviewed clinical experience as clinical data, loosens equivalence conditions, and lets bench, in vitro, in silico and simulation methods support safety and performance claims. The clinical evaluation consultation procedure would apply only to Class III implantables. The draft report pulls back by limiting expanded equivalence for high-risk implantables and strengthening expert panels.

  • Gains: Manufacturers whose EU strategy rests on equivalence to an established device, a common route for cost-competitive followers.
  • Loses or worries: German statutory insurers (DSV) oppose greater reliance on non-clinical data. The Biomedical Alliance warns against easier equivalence in clinical evaluation.

Odds of surviving trilogue: Medium. Expect the flexibility to survive for lower-risk classes and narrow sharply for implantables.

 

For exporters outside the EU: if your portfolio includes implantables, assume equivalence stays hard and keep clinical investigation plans live; for lower-risk lines, invest in strong literature and bench files.

6. Rule 11 and the software down-shift

The revised Rule 11 would let many software products fall into Class I instead of IIa, removing notified body involvement, with a new "serious" or "critical" situation test triggering higher classes. Baker McKenzie warns the wording leaves significant room for interpretation, and Mondaq's review of the draft report finds Parliament's changes do little more to ease Class I for software.

  • Gains: Developers of lower-risk clinical software. MedTech Europe wants Rule 11 aligned with international classification and Class I expanded further.
  • Loses or worries: DSV wants a notified body involved whenever software supports diagnostic or therapeutic decisions, AI at Class IIb or above, and self-learning systems at Class III.

On AI itself, briefly: the proposal moves the MDR and IVDR from Section A to Section B of Annex I to the AI Act, so AI devices would not face a parallel AI Act conformity assessment. The Council report shows many delegations supportive, some wanting sectoral AI rules; MedTech Europe asks for a delegated act within 12 months; ESIP wants high-risk AI requirements kept.

Odds of surviving trilogue: Medium. Down-classing survives in some form; the "decision support" boundary is where Parliament and Council will bargain.

 

For exporters outside the EU: India's health software exporters should classify under both the current and proposed Rule 11 and avoid launching on the assumption of Class I.

7. Fee cuts for micro and small firms

The proposal mandates notified body fee reductions of 50% for micro enterprises, 25% for small enterprises and 50% for orphan devices, and empowers the Commission to set fee levels and structures.

  • Gains: Small manufacturers, for whom certification costs bite hardest.
  • Loses or worries: The Council's June report warns the cuts could threaten notified bodies' financial sustainability, especially smaller ones, and questions the Commission's power over fees. Team-NB already reports that smaller notified bodies could face bankruptcy.

Odds of surviving trilogue: Medium. Some relief will pass; the size of the cut and who sets fees remain open.

 

For exporters outside the EU: watch this one closely. At Coreper on 14 October, one member state, backed by a few others, floated limiting simplification to EU-established SMEs; Asian small firms should press for neutral wording through their trade bodies now.

8. A route for well-established technologies

The proposal defines "well-established technology" devices (new Article 2(72) MDR) and drops the summary of safety and clinical performance and notified body review of periodic safety reports for qualifying higher-class devices. Separately, delegated regulations 2026/1359 and 2026/1451 have already extended the existing lists.

  • Gains: Makers of mature, high-volume product lines with long safety records.
  • Loses or worries: ESIP wants the category narrowed to low-risk, well-defined devices; Baker McKenzie flags vague criteria such as "simple" and "stable".

Odds of surviving trilogue: High. The Council supports the definition; the fight is over criteria.

 

For exporters outside the EU: assemble post-market data on long-running products now, since a strong history is the entry ticket once criteria are fixed.

9. Orphan, breakthrough and niche pathways

The proposal creates designated pathways, granted by expert panels, for orphan devices (conditions affecting no more than 12,000 people in the EU a year) and breakthrough devices, with priority and rolling review. The draft report adds a "niche" category for low-volume devices without suitable alternatives.

  • Gains: Specialist innovators and patients with few options. The Council discussion favours extending pathways to paediatric devices.
  • Loses or worries: Few object outright. DSV worries niche and orphan categories overlap; ESIP wants stronger post-market follow-up.

Odds of surviving trilogue: High. Broad support makes this the safest entry on the ledger.

 

For exporters outside the EU: Asian developers of rare-disease or paediatric devices should prepare designation dossiers early.

10. Looser in-house rules for laboratories

Last, a change with an unusual twist. The proposal drops the IVDR condition that no equivalent device exists before a hospital makes its own test, allows transfer between institutions, and brings EU-based central laboratories for clinical trials into scope. The draft report extends that to laboratories outside the EU serving clinical trials.

  • Gains: Hospital laboratories; the Biomedical Alliance welcomes the change.
  • Loses or worries: MedTech Europe wants the justification requirement in Article 5(5)(d) restored, and Emergo reports industry fears of direct competition.

Odds of surviving trilogue: Medium. Transfer rules look safe; removing the equivalence test is contested.

 

For exporters outside the EU: IVD kit sellers face softer hospital demand, while Asian central labs running EU trials could gain; follow the non-EU laboratory amendment.

◆   ◆   ◆

What the ledger tells us

Read the table from top to bottom and a pattern emerges. The broadest gains go to manufacturers who are already inside the system: holders of certificates, owners of clean audit histories, makers of mature product lines. The revision rewards incumbency far more than it rewards waiting.

That matters because, as Team-NB reports, some manufacturers are delaying applications in the expectation that requirements will fall. It is a risky bet. The current MDR and IVDR remain fully applicable, adoption is not expected before summer 2027, and even then the revised rules will not apply on day one.

For exporters in India and across Asia-Pacific, two quieter fights deserve more attention than the headline changes. The first is the international reliance chapter, which could, over time, let audits by trusted regulators count in Europe. The second is the SME question raised at Coreper, where a proposal to confine relief to EU-established firms would turn a simplification package into a competitive advantage for local players. Neither will be settled until trilogues begin in 2027.

The practical advice is unglamorous. Keep transition work moving, document everything that a risk-based reviewer could later reward, and make sure your trade association is speaking in Brussels before the Council text closes in December.

Organisations named in this ledger are welcome to respond; corrections of fact will be published.